Projects per year
Abstract
An improved miniature biosignal data sensor and recorder device is described, (NAT-1-4G) with 3-axis accelerometer, and a 2K Sa/sec all-channel recording capacity of 24 hours or more with a single zinc-air battery cell. Like the previous NAT-1 prototype device, this measures less than 18x22x10mm and weighs less than 2.3 grams, including the battery. In this paper we describe the device in detail, and introduce the presentation of tremor data measurement captured in the context of Parkinson’s disease fore-arm monitoring. The NAT-1-4G device has already achieved translation to commercialization and we expect it to become commercially available in the near future.
Original language | English |
---|---|
Pages | 84-92 |
Number of pages | 9 |
Publication status | Published - 16 Nov 2014 |
Event | Sensordevices 2014 : The Fifth International Conference on Sensor Device Technologies and Applications - lisbon, Portugal Duration: 16 Nov 2014 → 20 Nov 2014 |
Conference
Conference | Sensordevices 2014 : The Fifth International Conference on Sensor Device Technologies and Applications |
---|---|
Country/Territory | Portugal |
City | lisbon |
Period | 16/11/14 → 20/11/14 |
Projects
- 1 Active
-
NAT: NAT - Neural Acquisition Tracker
Crispin-Bailey, C., Austin, J., Moulds, A. & Freeman, M.
1/01/11 → …
Project: Other project › Research collaboration
File