Abstract
An intercomparison of three Fourier transform spectrometers (FTSs) with significantly different resolutions is presented. The highest-resolution instrument has a maximum optical path difference of 250 cm, and the two lower-resolution instruments have maximum optical path differences of 50 cm and 25 cm. The results indicate that the two lower-resolution instruments can retrieve total column amounts of O-3, HCl, N2O and CH4 using the SFIT2 retrieval code with percent differences from the high-resolution instrument generally better than 4%. Total column amounts of the stratospheric species (O-3 and HCl) have larger differences than those of the tropospheric species (N2O and CH4). Instrument line shape (ILS) information is found to be of critical importance when retrieving total columns of stratospheric gases from the lower-resolution instruments. Including the ILS information in the retrievals significantly reduces the difference in total column amounts between the three instruments. The remaining errors for stratospheric species total column amounts can be attributed to the lower sensitivity of the lower-resolution FTSs to the stratosphere.
Original language | English |
---|---|
Pages (from-to) | 1275-1292 |
Number of pages | 18 |
Journal | Atmospheric Chemistry and Physics |
Volume | 7 |
Publication status | Published - 22 Feb 2007 |
Keywords
- INFRARED MEASUREMENTS
- FTIR MEASUREMENTS
- LINE PARAMETERS
- HITRAN